24-hour consultation hotline:
86-15318987395
Products
Chlorophyll Fluorescence Imager
I. Product Overview
The Chlorophyll Fluorescence Imaging System is a professional instrument for plant photosynthesis research. Utilizing a high-sensitivity CMOS camera and LED light source control, it enables rapid, non-destructive detection and imaging analysis of chlorophyll fluorescence in plant leaves. The system integrates two core functions: OJIP rapid fluorescence kinetics measurement and PAM modulated fluorescence measurement, enabling comprehensive evaluation of plant photosynthetic system activity, light use efficiency, and responses to environmental stress.
Featuring a modular design with highly integrated hardware and software, the system offers a user-friendly graphical interface supporting bilingual Chinese-English switching. It is widely applicable for research and teaching in plant physiology, ecology, agricultural science, environmental science, and related fields.
II. Technical Specifications
Camera:
Resolution: 1608(H) × 1104(V)
Pixel Size: 9µm × 9µm
★Frame Rate: 100 fps
Pixel Depth: 12 bit
Interface Type: USB 3.0
Lens and Optical Parameters
★Focal Length: 12mm
★Maximum Aperture: F/2.8 (dynamically adjustable)
Horizontal Field of View (HFOV): Approx. 62.11°
Vertical Field of View (VFOV): Approx. 44.83°
Maximum Imaging Range: 50cm × 35cm at standard working distance
Light Source Parameters
Blue Light Wavelength: 450 nm excitation source for fluorescence excitation
Red Light Wavelength: 630 nm photochemical light for PAM measurement
Far-Infrared Light: 730 nm for specific measurement requirements
LED Brightness Range: 1% - 100%, peak intensity up to 1440 µmol/(m²·s)
Measurement Range
1. OJIP Measurement
Measurement Duration: 0.1 - 1.0 seconds, adjustable, recommended 1 second
LED Brightness: 1% - 100%, corresponding to 90-1440 µmol/(m²·s), 5% increments
2. PAM Measurement
Dark Adaptation Time: 0 - 3600 seconds, can be set to 0 to skip
Light Cycle Count: 1 - 100 cycles, number of cycles during light adaptation phase
Monitoring Duration: 10 - 120 seconds/cycle, monitoring time per cycle
Recovery Time: 60 - 600 seconds, dark recovery monitoring duration
Photochemical Light Intensity: 1000 - 30000, red light 630nm LED brightness value
Saturation pulse (Fm): 15%–100%, dark adaptation saturation pulse (450nm blue light)
Saturation pulse (Fm'): 15%–100%, light adaptation saturation pulse (450nm blue light)
Sampling Interval: 0.1 - 10.0 seconds, data acquisition interval
Measurement Accuracy
1. Fluorescence Intensity Measurement
Dynamic Range: 12-bit (0-4095)
Signal-to-Noise Ratio: >100:1
Repeatability: CV < 3%
Linearity: R² > 0.999
Sensitivity: Capable of detecting faint fluorescence signals
2. Parameter Calculation Accuracy
Fv/Fm: ±0.005
ΦPSII: ±0.01
qP/qN: ±0.02
NPQ: ±0.1
PIABS: ±0.05
3. Time Resolution
OJIP Mode: Minimum acquisition interval 10 ms